Onder Construction: Calibrating CV1 input of the mixers
At this page you find some info concerning the Fénix. If you are not sure please contact me by mail.
Why needs to calibrate. The CV1 input of the mixers are set for a gain of 1.00. There is probably no easy way to detect that you must
recalibrate this input. You can measure it.
How difficult is calibrating the CV1 input This is quit easy, you need a voltage meter to do this, Each CV1 Mixer input has his own trimpot.
Checking and Adjusting the CV1 input Put a voltage meter between the CV1 input and the output. Measure and memorize the reading.
Connect a voltage (Fi the output of the CPR) to the CV1 input.
The change in reading may max 5mV. Fi: without no voltage at the CV1 input you read 105 mV
and with a voltage to CV1 connected you read 108 mv. The change is 108-105= 3 mV. If you choose to recalibrate the only thing you do is adjust the trimpot so that the reading in the case that CV mixer 1 is
connected is close to the original reading.